Recall Z-0303-2017— Handicare USA, Inc.

Class II · Terminated

Class II FDA medical device recall by Handicare USA, Inc., initiated 2014-04-15, terminated 2016-12-13. FDA's recall record names no specific 510(k) clearance for this event. Product: ReTurn 7500/7500 Aid, transfer. Reason: During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder..

Recalling firm
Handicare USA, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-04-15
Terminated
2016-12-13
Firm location
Allentown, PA, United States

Product description

ReTurn 7500/7500 Aid, transfer

Reason for recall

During the period of April 2009  May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.

Data sourced from openFDA. This site is unofficial and independent of the FDA.