Recall Z-0301-2021— Smith & Nephew, Inc.

Class II · Terminated

Class II FDA medical device recall by Smith & Nephew, Inc., initiated 2020-09-21, terminated 2024-03-22. FDA's recall record names no specific 510(k) clearance for this event. Product: 5.5MM Arthrogarde Hip Access Cannula. Reason: The J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or arthroscope in the correct orientation..

Recalling firm
Smith & Nephew, Inc.
Classification
Class II
Status
Terminated
Initiated
2020-09-21
Terminated
2024-03-22
Firm location
Andover, MA, United States

Product description

5.5MM Arthrogarde Hip Access Cannula

Reason for recall

The J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or arthroscope in the correct orientation.