Recall Z-0295-2025— Philips Respironics, Inc.
Class I · Ongoing
Class I FDA medical device recall by Philips Respironics, Inc., initiated 2024-10-07. FDA's recall record names no specific 510(k) clearance for this event. Product: Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.. Reason: Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements..
- Recalling firm
- Philips Respironics, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2024-10-07
- Firm location
- Murrysville, PA, United States
Product description
Garbin Evo. Model Number: LD2110X23B. Intended for invasive and non-invasive ventilation of pediatric through adult patients weighing at least 2.5 kg in institutional, home, and non-emergency transport settings.
Reason for recall
Potential for in-line nebulizer configuration in certain locations to result in aerosol deposit accumulation on the device's internal flow sensor, which could lead to inaccurate flow measurements.