Recall Z-0294-2022— Medtronic Neuromodulation
Class II · Ongoing
Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2021-11-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.. Reason: The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed..
- Recalling firm
- Medtronic Neuromodulation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-11-04
- Firm location
- Minneapolis, MN, United States
Product description
Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.
Reason for recall
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Data sourced from openFDA. This site is unofficial and independent of the FDA.