Recall Z-0294-2013— MOOG Medical Devices Group
Class II · Terminated
Class II FDA medical device recall by MOOG Medical Devices Group, initiated 2012-11-08, terminated 2014-01-30. It names one FDA 510(k) clearance: K031199. Product: EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG. The device is used with disposable PVC administration tubing sets to provide a means for the volumetric delivery of fluids used in enteral applications. The EnteraLite Infinity enteral feeding pump with disposable PVC administration tubing sets and user accessories provide a means for the volumetric delivery of fluids used in enteral applications.. Reason: MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely..
- Recalling firm
- MOOG Medical Devices Group
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2012-11-08
- Terminated
- 2014-01-30
- Firm location
- Salt Lake City, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG. The device is used with disposable PVC administration tubing sets to provide a means for the volumetric delivery of fluids used in enteral applications. The EnteraLite Infinity enteral feeding pump with disposable PVC administration tubing sets and user accessories provide a means for the volumetric delivery of fluids used in enteral applications.
Reason for recall
MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.