Recall Z-0290-2019— Stryker GmbH
Class II · Terminated
Class II FDA medical device recall by Stryker GmbH, initiated 2018-08-28, terminated 2021-02-04. It names one FDA 510(k) clearance: K163656. Product: Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity Correction of segmental bony or soft tissue defects Joint arthrodesis Management of communicated intra-articular fractures of the distal radius Bone transport The Hoffmann LRF System is indicated in adults for: Osteotomy Revision procedure where other treatments or devices have been unsuccessful Bone reconstruction procedures Fusions and replantations of the foot Charcot foot reconstruction Lisfranc dislocations The typical LRF construct features two or more rings bridged by three to four telescopic struts with multiplanar, ball jointed fixation bolts. Once applied, the LRF can be unlocked for gross frame adjustment and fracture reduction. The LRF can be finely adjusted to aid in limb alignment and definitively locked to hold compression and reduction. If deemed appropriate, the LRF Telescopic Struts can provide up to 5mm of controlled dynamization. Original Statement OP-Tech H-ST-1_Rev2_HoffmannLRF_OpTech_2016-10690]. Reason: The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured..
- Recalling firm
- Stryker GmbH
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-28
- Terminated
- 2021-02-04
- Firm location
- Selzach, N/A, Switzerland
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Hoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment and fixation of: Open and closed fractures Post-traumatic joint contracture which has resulted in loss of range of motion Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction Pseudoarthrosis or non-union of long bones Limb lengthening by epiphyseal, diaphyseal, or metaphyseal distraction Correction of bony or soft tissue deformity Correction of segmental bony or soft tissue defects Joint arthrodesis Management of communicated intra-articular fractures of the distal radius Bone transport The Hoffmann LRF System is indicated in adults for: Osteotomy Revision procedure where other treatments or devices have been unsuccessful Bone reconstruction procedures Fusions and replantations of the foot Charcot foot reconstruction Lisfranc dislocations The typical LRF construct features two or more rings bridged by three to four telescopic struts with multiplanar, ball jointed fixation bolts. Once applied, the LRF can be unlocked for gross frame adjustment and fracture reduction. The LRF can be finely adjusted to aid in limb alignment and definitively locked to hold compression and reduction. If deemed appropriate, the LRF Telescopic Struts can provide up to 5mm of controlled dynamization. Original Statement OP-Tech H-ST-1_Rev2_HoffmannLRF_OpTech_2016-10690]
Reason for recall
The press fit connection of the Angular Sleeve and pin of the U-Joint and Click Lock did not meet specification. A dimensional discrepancy of the bore hole of the Angular Sleeve may cause a reduced press fit of the pin. This can result in partial of full displacement or loss of the pin and, accordingly, stability of the strut of the frame cannot be ensured.
Data sourced from openFDA. This site is unofficial and independent of the FDA.