Recall Z-0288-2024— Olympus Corporation of the Americas

Class II · Ongoing

Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2023-10-02. It names one FDA 510(k) clearance: K112680. Product: EVIS EXERA III Gastrointestinal Videoscope. Reason: Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel..

Recalling firm
Olympus Corporation of the Americas
Classification
Class II
Status
Ongoing
Initiated
2023-10-02
Firm location
Center Valley, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

EVIS EXERA III Gastrointestinal Videoscope

Reason for recall

Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.