Recall Z-0288-2024— Olympus Corporation of the Americas
Class II · Ongoing
Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2023-10-02. It names one FDA 510(k) clearance: K112680. Product: EVIS EXERA III Gastrointestinal Videoscope. Reason: Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel..
- Recalling firm
- Olympus Corporation of the Americas
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-10-02
- Firm location
- Center Valley, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EVIS EXERA III Gastrointestinal Videoscope
Reason for recall
Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.