Recall Z-0286-2020— Biomedical Polymers, Inc.

Class III · Terminated

Class III FDA medical device recall by Biomedical Polymers, Inc., initiated 2019-09-18, terminated 2020-09-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Hemo-Drop Blood Dispenser, Part Number BMP-HEMODROP, used for preparing blood samples.. Reason: The metal cannula of the device either broke and the broken piece left in the blood specimen tube, or slid back into the plastic portion of the device instead of piercing the tube..

Recalling firm
Biomedical Polymers, Inc.
Classification
Class III
Status
Terminated
Initiated
2019-09-18
Terminated
2020-09-15
Firm location
Sterling, MA, United States

Product description

Hemo-Drop Blood Dispenser, Part Number BMP-HEMODROP, used for preparing blood samples.

Reason for recall

The metal cannula of the device either broke and the broken piece left in the blood specimen tube, or slid back into the plastic portion of the device instead of piercing the tube.

Data sourced from openFDA. This site is unofficial and independent of the FDA.