Recall Z-0286-2013— Stryker Howmedica Osteonics Corp.

Class II · Terminated

Class II FDA medical device recall by Stryker Howmedica Osteonics Corp., initiated 2012-07-16, terminated 2015-09-09. FDA's recall record names no specific 510(k) clearance for this event. Product: Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.. Reason: Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem..

Recalling firm
Stryker Howmedica Osteonics Corp.
Classification
Class II
Status
Terminated
Initiated
2012-07-16
Terminated
2015-09-09
Firm location
Mahwah, NJ, United States

Product description

Special Orthopaedic Solution Peri-Acetabular Prosthesis V40 Femoral Extension Component Product Usage: The PAR extension piece is a component of the PAR assembly. The Peri-Acetabular Reconstruction (PAR) endoprosthesis was developed to meet the unique needs of patients who require reconstruction of large acetabular defects after a Type I and II pelvic resection.

Reason for recall

Stryker has become aware that insufficient taper torsional strength may result when a PAR extension piece is used with a forged CoCr stem.