Recall Z-0285-2025— Zimmer, Inc.

Class II · Ongoing

Class II FDA medical device recall by Zimmer, Inc., initiated 2024-09-17. It names one FDA 510(k) clearance: K140603. Product: Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273). Reason: A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery.

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-09-17
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Saphena Medical Venapax XL- Endoscopic Vessel Harvesting System for use in minimally invasive surgery allowing access for vessel harvesting and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. Item Number: VPX4000 (7273)

Reason for recall

A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery