Recall Z-0283-2021— Intuitive Surgical, Inc.
Class II · Terminated
Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2019-06-17, terminated 2022-07-18. It names one FDA 510(k) clearance: K131861. Product: da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.. Reason: One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images..
- Recalling firm
- Intuitive Surgical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-06-17
- Terminated
- 2022-07-18
- Firm location
- Sunnyvale, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Reason for recall
One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
Data sourced from openFDA. This site is unofficial and independent of the FDA.