Recall Z-0281-2019— Leica Microsystems, Inc.

Class II · Terminated

Class II FDA medical device recall by Leica Microsystems, Inc., initiated 2018-09-17, terminated 2020-08-21. FDA's recall record names no specific 510(k) clearance for this event. Product: BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments. Reason: The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound..

Recalling firm
Leica Microsystems, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-09-17
Terminated
2020-08-21
Firm location
Buffalo Grove, IL, United States

Product description

BOND Slide Labeler Zebra Printer Model TLP3842 and Model GX4301, for use with BOND-III and BOND-MAX instruments

Reason for recall

The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.