Recall Z-0281-2014— RAYSEARCH LABORATORIES AB

Class II · Terminated

Class II FDA medical device recall by RAYSEARCH LABORATORIES AB, initiated 2013-10-23, terminated 2017-02-17. It names 2 FDA 510(k) clearances: K130617, K120387. Product: Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0. Reason: Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees..

Recalling firm
RAYSEARCH LABORATORIES AB
Classification
Class II
Status
Terminated
Initiated
2013-10-23
Terminated
2017-02-17
Firm location
Stockholm, Sweden

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ray Station Radiation Therapy Treatment Planning System, Model Numbers 3.0, 3.5 and 4.0

Reason for recall

Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.