Recall Z-0278-2014— Medtronic Neuromodulation
Class II · Terminated
Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2013-09-30, terminated 2015-11-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.. Reason: Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR)..
- Recalling firm
- Medtronic Neuromodulation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-09-30
- Terminated
- 2015-11-20
- Firm location
- Minneapolis, MN, United States
Product description
Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
Reason for recall
Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).