Recall Z-0278-2014— Medtronic Neuromodulation

Class II · Terminated

Class II FDA medical device recall by Medtronic Neuromodulation, initiated 2013-09-30, terminated 2015-11-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.. Reason: Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR)..

Recalling firm
Medtronic Neuromodulation
Classification
Class II
Status
Terminated
Initiated
2013-09-30
Terminated
2015-11-20
Firm location
Minneapolis, MN, United States

Product description

Medtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.

Reason for recall

Medtronic notified customers regarding Loss of Stimulation of specific Implantable Neuro-Stimulators under the following specific conditions: Switching between groups with Multiple programs or creating program groups. Therapy can be restored with the Physician Recharge mode (RPM) of the Implantable Neurostimulator Recharger (INSR).

Data sourced from openFDA. This site is unofficial and independent of the FDA.