Recall Z-0276-2019— Medtronic Navigation, Inc.-Littleton
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc.-Littleton, initiated 2018-08-30, terminated 2020-06-18. It names one FDA 510(k) clearance: K092564. Product: O-arm 1000 2nd Edition Imaging System, BASE BASE OARM Bl70000027120R SYSTEM 120V RWK, Model Number Bl70000027120R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.. Reason: Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems..
- Recalling firm
- Medtronic Navigation, Inc.-Littleton
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-30
- Terminated
- 2020-06-18
- Firm location
- Littleton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
O-arm 1000 2nd Edition Imaging System, BASE BASE OARM Bl70000027120R SYSTEM 120V RWK, Model Number Bl70000027120R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Reason for recall
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.