Recall Z-0274-2019— Medtronic Navigation, Inc.-Littleton
Class II · Terminated
Class II FDA medical device recall by Medtronic Navigation, Inc.-Littleton, initiated 2018-08-30, terminated 2020-06-18. It names one FDA 510(k) clearance: K092564. Product: O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100R SYSTEM lOOV RWK, Model Number Bl70000027100R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.. Reason: Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems..
- Recalling firm
- Medtronic Navigation, Inc.-Littleton
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-30
- Terminated
- 2020-06-18
- Firm location
- Littleton, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027100R SYSTEM lOOV RWK, Model Number Bl70000027100R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Reason for recall
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
Data sourced from openFDA. This site is unofficial and independent of the FDA.