Recall Z-0273-2016— Philips Electronics North America Corporation
Class II · Terminated
Class II FDA medical device recall by Philips Electronics North America Corporation, initiated 2014-11-19, terminated 2017-09-12. It names 4 FDA 510(k) clearances: K061707, K051134, K040404 and 1 more. Product: ECG Out Cable,for use with a Philips ALS monitor/ defibrillator and bedside monitor for Synchronized cardioversion and Transcutaneous pacing. M1783A/M5526A 12-pin Sync cables, used with HeartStart MRx and HeartStart XL and HeartStart XL+. Reason: When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs) while connected to AC power: On the HeartStart MRx and HeartStart XL, EFT noise can be misinterpreted as an R-wave. On the HeartStart XL+, EFT noise can disable ECG monitoring, and potentially interrupt demand mode pacing..
- Recalling firm
- Philips Electronics North America Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-11-19
- Terminated
- 2017-09-12
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
4 on FDA's record- K061707 — HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
- K051134 — HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
- K040404 — HEARTSTART MRX, MODEL M3535A
- K031187 — HEARTSTART MRX, MODEL M3535A
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
ECG Out Cable,for use with a Philips ALS monitor/ defibrillator and bedside monitor for Synchronized cardioversion and Transcutaneous pacing. M1783A/M5526A 12-pin Sync cables, used with HeartStart MRx and HeartStart XL and HeartStart XL+
Reason for recall
When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs) while connected to AC power: On the HeartStart MRx and HeartStart XL, EFT noise can be misinterpreted as an R-wave. On the HeartStart XL+, EFT noise can disable ECG monitoring, and potentially interrupt demand mode pacing.