Recall Z-0271-2013— Sorin Group USA, Inc.

Class II · Terminated

Class II FDA medical device recall by Sorin Group USA, Inc., initiated 2012-09-19, terminated 2013-01-18. It names 2 FDA 510(k) clearances: K093882, K071318. Product: S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.. Reason: Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi.

Recalling firm
Sorin Group USA, Inc.
Classification
Class II
Status
Terminated
Initiated
2012-09-19
Terminated
2013-01-18
Firm location
Arvada, CO, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

S5 Single Roller Pump 150, Item Number: 10-80-00 The Stockert S5/Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six hours or less.

Reason for recall

Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Sorin Group S5 Perfusion System due to reports of a "Fault in motor controller" alarm message displayed on the system panel. This could lead to unintended pump stoppage which may result in permanent impairment or life-threatening injury. An out of specification electrical component was determi

Data sourced from openFDA. This site is unofficial and independent of the FDA.