Recall Z-0266-2018— Medacta Usa Inc
Class II · Terminated
Class II FDA medical device recall by Medacta Usa Inc, initiated 2017-10-24, terminated 2018-06-20. It names one FDA 510(k) clearance: K090988. Product: The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF. Reason: Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques..
- Recalling firm
- Medacta Usa Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-10-24
- Terminated
- 2018-06-20
- Firm location
- Memphis, TN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The GMK knee system: designed for cemented use in total knee arthroplasty. Sphere and Primary Inserts with part number ranges: 02.12.0110FL - 02.12.0620FL; 02.12.0110FR - 02.12.0620FR; 02.07.0110PSF - 02.07.0620PSF
Reason for recall
Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.