Recall Z-0254-2016— GE Healthcare
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, initiated 2015-08-31, terminated 2018-09-07. It names one FDA 510(k) clearance: K132376. Product: GE Healthcare, Discovery MR750w 3.0T. Product Usage: The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T Systems are whole body magnetic resonance scanners for diagnostic imaging.. Reason: When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans..
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-08-31
- Terminated
- 2018-09-07
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare, Discovery MR750w 3.0T. Product Usage: The Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T and the Optima MR450w 1.5T Systems are whole body magnetic resonance scanners for diagnostic imaging.
Reason for recall
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Data sourced from openFDA. This site is unofficial and independent of the FDA.