Recall Z-0247-2016— Maquet Cardiovascular Us Sales, Llc
Class II · Terminated
Class II FDA medical device recall by Maquet Cardiovascular Us Sales, Llc, initiated 2015-08-17, terminated 2016-10-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.. Reason: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events..
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-08-17
- Terminated
- 2016-10-19
- Firm location
- Wayne, NJ, United States
Product description
Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.
Reason for recall
Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.
Data sourced from openFDA. This site is unofficial and independent of the FDA.