Recall Z-0247-2014— Carestream Health Inc.
Class II · Terminated
Class II FDA medical device recall by Carestream Health Inc., initiated 2013-09-16, terminated 2015-12-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Carestream Vue PACS; MX WEB SW/ 1 MX-O CONCUR ACCESS LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.. Reason: Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem..
- Recalling firm
- Carestream Health Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-09-16
- Terminated
- 2015-12-21
- Firm location
- Rochester, NY, United States
Product description
Carestream Vue PACS; MX WEB SW/ 1 MX-O CONCUR ACCESS LIC The CARESTREAM PACS is an image management system whose intended use is to provide completely scaleable local and wide area PACS solutions for hospital and related institutions/sites, which will archive, distribute, retrieve and display images and data from all hospital modalities and information systems.
Reason for recall
Carestream Health, Inc. has issued a recall when using the CARESTREAM Client Workstation where under very specific conditions, some images from multi-slice studies such as CT and MR may not load into the viewer application due to a software problem.
Data sourced from openFDA. This site is unofficial and independent of the FDA.