Recall Z-0245-2018— Angiodynamics Inc. (Navilyst Medical Inc.)
Class II · Terminated
Class II FDA medical device recall by Angiodynamics Inc. (Navilyst Medical Inc.), initiated 2016-11-28, terminated 2019-06-12. It names one FDA 510(k) clearance: K150448. Product: Bio-Stable 5F DL-55CM IR-145 Kit Valved PG, UPN H965458330, Catalog No. 45-833 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.. Reason: Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits..
- Recalling firm
- Angiodynamics Inc. (Navilyst Medical Inc.)
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-11-28
- Terminated
- 2019-06-12
- Firm location
- Glens Falls, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Bio-Stable 5F DL-55CM IR-145 Kit Valved PG, UPN H965458330, Catalog No. 45-833 The Xcela with PASV, BioFlo PICC with ENDEXO Technology, and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring.
Reason for recall
Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
Data sourced from openFDA. This site is unofficial and independent of the FDA.