Recall Z-0242-2022— GE Healthcare, LLC
Class II · Ongoing
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2021-06-30. It names 3 FDA 510(k) clearances: K191777, K163213, K133705. Product: Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition. Reason: The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered..
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-06-30
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
3 on FDA's record- K191777 — Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
- K163213 — Revolution CT
- K133705 — REVOLUTION CT
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Computed Tomography Systems Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Reason for recall
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.