Recall Z-0241-2023— Steris Corporation

Class II · Terminated

Class II FDA medical device recall by Steris Corporation, initiated 2022-10-17, terminated 2024-05-28. It names one FDA 510(k) clearance: K192265. Product: GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001. Reason: Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result.

Recalling firm
Steris Corporation
Classification
Class II
Status
Terminated
Initiated
2022-10-17
Terminated
2024-05-28
Firm location
Mentor, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Reason for recall

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result