Recall Z-0238-2022— Philips Healthcare
Class II · Ongoing
Class II FDA medical device recall by Philips Healthcare, initiated 2021-09-24. It names one FDA 510(k) clearance: K200917. Product: Philips Azurion systems with software releases 2.1(L1) and 2.1(L2). Reason: 3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software..
- Recalling firm
- Philips Healthcare
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-09-24
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips Azurion systems with software releases 2.1(L1) and 2.1(L2)
Reason for recall
3D-RA is a reconstruction software product that can be used with the Philips Azurin system. This product extends the functionality of compatible X-ray equipment with the capability of 3D X-ray imaging. Before starting a 3D-RA scan, the detector of the Philips Azurion system must be positioned in portrait or landscape orientation with an angle of 0 or +/-90 degrees relative to the 3D scan direction. Philips has discovered that Azurion releases 2.1(L1) and 2.1(L2) allow creation of a 3D-RA scan with a deviating detector orientation of up to 3 degrees from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees), relative to the 3D scan direction. 3D-RA scans with a deviating detector orientation of ore than 1.0 degree from the exact portrait or landscape orientation (i.e., at 0 or +/-90 degrees) cannot be reconstructed with the 3D-RA reconstruction software.
Data sourced from openFDA. This site is unofficial and independent of the FDA.