Recall Z-0233-2016— Utah Medical Products, Inc
Class II · Terminated
Class II FDA medical device recall by Utah Medical Products, Inc, initiated 2015-10-16, terminated 2016-05-02. It names one FDA 510(k) clearance: K821814. Product: Myelo-Nate CSF Sampling Kit, Neonatal/Pediatric cerebralspinal fluid sampling sterile tray, including a lumbar puncture needle, four collection vials, an iodine swabstick package, a fenestrated drape, a polylined drape, three gauze pads, and one spot bandage.. Reason: Utah Medical is conducting a recall of Myelo-Nate CSF Sampling Kit due to an internal post-sterilization inspection that found a sterile finished tray where the seal between the Tyvek lid and the PETG tray on one side of the tray had been compromised..
- Recalling firm
- Utah Medical Products, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-10-16
- Terminated
- 2016-05-02
- Firm location
- Midvale, UT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Myelo-Nate CSF Sampling Kit, Neonatal/Pediatric cerebralspinal fluid sampling sterile tray, including a lumbar puncture needle, four collection vials, an iodine swabstick package, a fenestrated drape, a polylined drape, three gauze pads, and one spot bandage.
Reason for recall
Utah Medical is conducting a recall of Myelo-Nate CSF Sampling Kit due to an internal post-sterilization inspection that found a sterile finished tray where the seal between the Tyvek lid and the PETG tray on one side of the tray had been compromised.