Recall Z-0232-2016— Boston Scientific Corporation

Class II · Terminated

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2015-09-17, terminated 2016-03-08. FDA's recall record names no specific 510(k) clearance for this event. Product: Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.. Reason: Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte.

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Terminated
Initiated
2015-09-17
Terminated
2016-03-08
Firm location
Saint Paul, MN, United States

Product description

Boston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.

Reason for recall

Devices shipped to US in dual-channel rather than US approved single-channel RF Telemetry communication mode. While this does not impact therapy delivery of device; this can impact likelihood of successfully establishing telemetry during first scan of device when using a single-channel programmer. Telemetry can be established with additional scans; once established no further difficulty is expecte