Recall Z-0229-2027— Boston Scientific Cork, Ltd.

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Cork, Ltd., initiated 2026-09-16. It names one FDA 510(k) clearance: K163701. Product: Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260. Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment..

Recalling firm
Boston Scientific Cork, Ltd.
Classification
Class II
Status
Ongoing
Initiated
2026-09-16
Firm location
Cork, Ireland

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260

Reason for recall

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Data sourced from openFDA. This site is unofficial and independent of the FDA.