Recall Z-0229-2026— Synthes (USA) Products LLC
Class II · Ongoing
Class II FDA medical device recall by Synthes (USA) Products LLC, initiated 2025-09-26. It names 2 FDA 510(k) clearances: K201346, K201336. Product: LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.. Reason: Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm..
- Recalling firm
- Synthes (USA) Products LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-09-26
- Firm location
- West Chester, PA, United States
510(k) clearances named in this recall
2 on FDA's record- K201346 — DePuy Synthes Retrograde Femoral Nail Advanced System, DePuy Synthes Locking Screws for Medullary Nails, 5.0 mm
- K201336 — DePuy Synthes Tibial Nail Advanced, Depuy Synthes Locking Screws Medullary Nails, 4.0 and 5.0mm
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LOW PROF LCKNG SCREW, F/IM NAIL 5.0MM / L, 42MM / XL25/ STER. Product Code: 04.045.342TS. Used in RFN-ADVANCEDTM Retrograde Femoral Nailing (RFNA) System and the Tibial Nail Advanced System.
Reason for recall
Although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.