Recall Z-0226-2023— Boston Scientific Corporation

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2022-10-10. It names one FDA 510(k) clearance: K171454. Product: ORISE ProKnife Procedure Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device. UPN (1)ORISE ProKnife 1.5 mm Electrode - Kit M00519380; (2) ORISE ProKnife 2.0 mm Electrode - Kit M00519390; (3) ORISE ProKnife 3.0 mm Electrode - Kit M00519400. Reason: ***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention.

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Ongoing
Initiated
2022-10-10
Firm location
Marlborough, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ORISE ProKnife Procedure Kits: indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device. UPN (1)ORISE ProKnife 1.5 mm Electrode - Kit M00519380; (2) ORISE ProKnife 2.0 mm Electrode - Kit M00519390; (3) ORISE ProKnife 3.0 mm Electrode - Kit M00519400

Reason for recall

***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention