Recall Z-0224-2023— Boston Scientific Corporation

Class II · Ongoing

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2022-10-10. It names one FDA 510(k) clearance: K180068. Product: ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device Packaged as follows: UPN: (1)Twin Pack Box 1 M00519200 (2)Twin Pack Box 10 M00519201 (3)Single Pack Box 1 M00519220 (4)Single Pack Box 10 M00519221. Reason: ***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention.

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Ongoing
Initiated
2022-10-10
Firm location
Marlborough, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ORISE Gel Submucosal Lifting Agent--Indicated for use in gastrointestinal endoscopic procedures for submucosal lift of polyps, adenomas, early stage cancers or other gastrointestinal mucosal lesions, prior to excision with a snare or other suitable endoscopic device Packaged as follows: UPN: (1)Twin Pack Box 1 M00519200 (2)Twin Pack Box 10 M00519201 (3)Single Pack Box 1 M00519220 (4)Single Pack Box 10 M00519221

Reason for recall

***Update12/13/22***:Product removal due to higher incidence of adverse events of foreign body reactions*** Potential outcomes associated with foreign body reaction to remnant ORlSE Gel post procedure. Foreign body reaction which presented as mass formations from remnant ORlSE Gel post-procedure, prompting surgical intervention