Recall Z-0220-2023— Ortho-Clinical Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Ortho-Clinical Diagnostics, Inc., initiated 2022-10-13. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT. Reason: Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results..

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2022-10-13
Firm location
Raritan, NJ, United States

Product description

VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT

Reason for recall

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.