Recall Z-0219-2024— Magellan Diagnostics, Inc.
Class II · Ongoing
Class II FDA medical device recall by Magellan Diagnostics, Inc., initiated 2023-09-19. It names one FDA 510(k) clearance: K052549. Product: LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.. Reason: During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result)..
- Recalling firm
- Magellan Diagnostics, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-09-19
- Firm location
- North Billerica, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Reason for recall
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).