Recall Z-0219-2016— Philips Medical Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems, Inc., initiated 2014-12-23, terminated 2020-09-09. It names 6 FDA 510(k) clearances: K063375, K062233, K061707 and 3 more. Product: HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. Reason: MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed.
- Recalling firm
- Philips Medical Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-12-23
- Terminated
- 2020-09-09
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
6 on FDA's record- K063375 — HEARTSTART MRX, MODELS M3535A OR M3536A
- K062233 — HEARTSTART MRX MONITOR/DEFIBRILLATOR
- K061707 — HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
- K051134 — HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
- K040404 — HEARTSTART MRX, MODEL M3535A
- K031187 — HEARTSTART MRX, MODEL M3535A
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Reason for recall
MRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed