Recall Z-0213-2015— Masimo Corporation
Class II · Terminated
Class II FDA medical device recall by Masimo Corporation, initiated 2014-09-30, terminated 2016-03-28. It names one FDA 510(k) clearance: K121914. Product: Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate when installed in a compatible OEM host system. Shipping/Package: All Pulse Oximetry cables are shipped non-sterile in a cardboard carton. They are intended to be used non-sterile.. Reason: Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device..
- Recalling firm
- Masimo Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-09-30
- Terminated
- 2016-03-28
- Firm location
- Irvine, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Masimo SET uSpO2 Cable The Masimo SET uSpO2 Pulse Oximetry Cable is an in-line cable that provides continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate when installed in a compatible OEM host system. Shipping/Package: All Pulse Oximetry cables are shipped non-sterile in a cardboard carton. They are intended to be used non-sterile.
Reason for recall
Firm has identified a small number of Oximetry Cables with crossed-internal wires. The effect on performance when these wires are crossed is that when a low SpO2 value is measured, a high SpO2 value would be displayed on the monitoring device. Additionally when a high SpO2 value is measured, a low SpO2 value would be displayed on the monitoring device.