Recall Z-0210-2018— Biosense Webster, Inc.

Class II · Terminated

Class II FDA medical device recall by Biosense Webster, Inc., initiated 2017-06-14, terminated 2021-01-11. FDA's recall record names no specific 510(k) clearance for this event. Product: SMARTABLATE RF System Generator. Reason: Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets..

Recalling firm
Biosense Webster, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-14
Terminated
2021-01-11
Firm location
Irwindale, CA, United States

Product description

SMARTABLATE RF System Generator

Reason for recall

Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.