Recall Z-0209-2014— Hospira Inc.

Class II · Terminated

Class II FDA medical device recall by Hospira Inc., initiated 2012-02-05, terminated 2016-11-14. It names one FDA 510(k) clearance: K042081. Product: *** 1) Plum A+ Infusion Pump; List Number: 20678; Labeled in part: PLUM A+3; for use with Hospira MedNet; *** 2) Plum A+ Infusion Pump; List Number: 20679; Labeled in part: PLUM A+; for use with Hospira MedNet; *** 3) Plum A+ Infusion Pump with Hospira MedNet Software; List Number: 20792; Labeled in part: PLUM A+; for use with Hospira MedNet; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. Reason: Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not.

Recalling firm
Hospira Inc.
Classification
Class II
Status
Terminated
Initiated
2012-02-05
Terminated
2016-11-14
Firm location
Lake Forest, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

*** 1) Plum A+ Infusion Pump; List Number: 20678; Labeled in part: PLUM A+3; for use with Hospira MedNet; *** 2) Plum A+ Infusion Pump; List Number: 20679; Labeled in part: PLUM A+; for use with Hospira MedNet; *** 3) Plum A+ Infusion Pump with Hospira MedNet Software; List Number: 20792; Labeled in part: PLUM A+; for use with Hospira MedNet; Product Usage: The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care

Reason for recall

Hospira has received reports of distal pressure sensor calibration drift on Plum A+ Infusers. The following error codes may be displayed during setup or infusion signifying possible calibration drift: E180/N180, E181/N181, E186/N186, E187/N187 or E346. Additionally, if the pressure sensor calibration has drifted, users may experience false distal occlusion alarms or distal occlusions that may not

Data sourced from openFDA. This site is unofficial and independent of the FDA.