Recall Z-0203-2019— Data Innovations, LLC
Class II · Terminated
Class II FDA medical device recall by Data Innovations, LLC, initiated 2018-08-29, terminated 2019-02-05. FDA's recall record names no specific 510(k) clearance for this event. Product: IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that can receive information from a hospital computer system (also called a Laboratory Information System [LIS]).. Reason: Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction..
- Recalling firm
- Data Innovations, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-29
- Terminated
- 2019-02-05
- Firm location
- South Burlington, VT, United States
Product description
IM Abbott Alinity ci-series (abdicqii) accessory driver v8.00.0001 software product that can receive information from a hospital computer system (also called a Laboratory Information System [LIS]).
Reason for recall
Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
Data sourced from openFDA. This site is unofficial and independent of the FDA.