Recall Z-0200-2023— Covidien, LP
Class II · Ongoing
Class II FDA medical device recall by Covidien, LP, initiated 2022-09-28. It names one FDA 510(k) clearance: K111442. Product: Endo Stitch V-Loc 180 Absorbable Reload, Models VLOCA004L, VLOCA006L, VLOCA008L, VLOCA204L, VLOCA206L, VLOCA208L, VLOCA304L, VLOCA306L, VLOCA308L. Automated suturing device.. Reason: Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios..
- Recalling firm
- Covidien, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-09-28
- Firm location
- North Haven, CT, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Endo Stitch V-Loc 180 Absorbable Reload, Models VLOCA004L, VLOCA006L, VLOCA008L, VLOCA204L, VLOCA206L, VLOCA208L, VLOCA304L, VLOCA306L, VLOCA308L. Automated suturing device.
Reason for recall
Medtronic has received 210 reports of needles breaking during endoscopic suturing. The potential harm(s) include a delay in treatment/therapy, the potential for a portion of the needle to remain in the patient, foreign body reaction, allergic reaction, tissue injury and unintended radiation exposure from additional imaging for both intra-op and postoperative scenarios.