Recall Z-0190-2014— Advanced Sterilization Products

Class II · Terminated

Class II FDA medical device recall by Advanced Sterilization Products, initiated 2013-10-18, terminated 2014-01-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Sterrad 100NX, Product Code: 10104. Used to sterilize heat and moisture sensitive reusable medical devices.. Reason: Advanced Sterilization Products (ASP) is recalling the Sterrad 100NX Sterilizer some healthcare workers detected odors/smells emanating from the 100NX Sterilizers, and it may be related to normal breakdown of the oil used in the Sterrad 200 vacuum pumps..

Recalling firm
Advanced Sterilization Products
Classification
Class II
Status
Terminated
Initiated
2013-10-18
Terminated
2014-01-16
Firm location
Irvine, CA, United States

Product description

Sterrad 100NX, Product Code: 10104. Used to sterilize heat and moisture sensitive reusable medical devices.

Reason for recall

Advanced Sterilization Products (ASP) is recalling the Sterrad 100NX Sterilizer some healthcare workers detected odors/smells emanating from the 100NX Sterilizers, and it may be related to normal breakdown of the oil used in the Sterrad 200 vacuum pumps.