Recall Z-0189-2020— Smiths Medical ASD Inc.
Class II · Terminated
Class II FDA medical device recall by Smiths Medical ASD Inc., initiated 2019-07-08, terminated 2020-08-23. It names one FDA 510(k) clearance: K870866. Product: GRIPPER Needles sold individually. Labeled as the following item descriptions: 1. NDL, GRIPPER, 22G X 1" (25MM), 12/BX; 2. NDL, GRIPPER, 22G X 1.25" (32MM), 12/BX; 3. NDL, GRIPPER, 20G X 1.25" (32MM), 12/BX; 4. NDL, GRIPPER, 19G X 1" (25MM), 12/BX; 5. NDL, GRIPPER, 19G X 1.25" (32MM), 12/BX; 6. NDL, GRIPPER, 22G X .75" (19MM), 12/BX; 7. NDL, GRIPPER, 19G X .75" (19MM), 12/BX; 8. NDL, GRIPPER, 20G X .625" (16MM), 12/BX; 9. NDL, GRIPPER, 22G X .625" (16MM), 12/BX; 10. NDL, GRIPPER PLUS, 22G X 1" (25MM), 12/BX; 11. NDL, GRIPPER PLUS, 22G X 1.25" (32MM), 12/BX; 12. NDL, GRIPPER PLUS, 20G X 1" (25MM) 12/BX; 13. NDL, GRIPPER PLUS, 20G X 1.25" (32MM) 12/BX; 14. NDL, GRIPPER PLUS, 19G X 1" (25MM) 12/BX; 15. NDL, GRIPPER PLUS, 19G X 1.25" (32MM) 12/BX; 16. NDL, GRIPPER PLUS, 22G X .75" (19MM) 12/BX; 17. NDL, GRIPPER PLUS, 20G X .75" (19MM) 12/BX; 18. NDL, GRIPPER PLUS, 19G X .75" (19MM) 12/BX; 19. NDL, GRIPPER PLUS, 20G X .625" (16MM) 12/BX; 20. NDL, GRIPPER PLUS, 22G X .625" (16MM) 12/BX; 21. NDL, GRIPPER PLUS, 22G X .625" (16MM), Y-SITE 12/BX; 22. NDL, GRIPPER PLUS, 22G X .75" (19MM), Y-SITE 12/BX; 23. NDL, GRIPPER PLUS, 22G X 1" (25MM), Y-SITE 12/BX; 24. NDL, GRIPPER PLUS, 22G X 1.25" (32MM), Y-SITE 12/BX; 25. NDL, GRIPPER PLUS, 20G X .625" (16MM), Y-SITE 12/BX; 26. NDL, GRIPPER PLUS, 20G X .75" (19MM), Y-SITE 12/BX; 27. NDL, GRIPPER PLUS, 20G X 1" (25MM), Y-SITE 12/BX; 28. NDL, GRIPPER PLUS, 20G X 1.25" (32MM), Y-SITE 12/BX; 29. NDL, GRIPPER PLUS, 19G X .75" (19MM), Y-SITE 12/BX; 30. NDL, GRIPPER PLUS, 19G X 1" (25MM), Y-SITE 12/BX; 31. NDL, GRIPPER PLUS, 19G X 1.25" (32MM), Y-SITE 12/BX; 32. NDL, GRIPPER, 22G X .625" (16MM), Y-SITE 12/BX; 33. NDL, GRIPPER, 22G X .75" (19MM), Y-SITE 12/BX; 34. NDL, GRIPPER, 22G X 1" (25MM), Y-SITE 12/BX; 35. NDL, GRIPPER, 22G X 1.25" (32MM), Y-SITE 12/BX; 36. NDL, GRIPPER, 20G X .625" (16MM), Y-SITE 12/BX; 37. NDL, GRIPPER, 20G X .75" (19MM), Y-SITE 12/BX; 38. NDL, GRIPPER, 20G X 1" (25MM), Y-SITE 12/BX; 39. NDL, GRIPPER, 20G X 1.25" (32MM), Y-SITE 12/BX; 40. NDL, GRIPPER, 19G X .75" (19MM), Y-SITE 12/BX; 41. NDL, GRIPPER, 19G X 1.25" (32MM), Y-SITE 12/BX; 42. NDL, GRIPPER PLUS, 22G X .625" (16MM), NEEDLELESS Y-SITE 12/BX; 43. NDL, GRIPPER PLUS, 22G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 44. NDL, GRIPPER PLUS, 22G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 45. NDL, GRIPPER PLUS, 20G X .625" (16MM), NEEDLESS Y-SITE 12/BX; 46. NDL, GRIPPER PLUS, 20G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 47. NDL, GRIPPER PLUS, 20G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 48. NDL, GRIPPER PLUS, 20G X 1.25" (32MM), NEEDLELESS Y-SITE 12/BX; 49. NDL, GRIPPER PLUS, 19G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 50. NDL, GRIPPER PLUS, 19G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 51. NDL, GRIPPER PLUS, 19G X 1.25" (32MM), NEEDLELESS Y-SITE 12/BX; 52. NDL, GRIPPER MICRO, 22G X .75 12/BOX; 53. NDL, GRIPPER MICRO, 22G X 1" 12/BOX; 54. NDL, GRIPPER MICRO, 22G X 1.25" 12/BOX; 55. NDL, GRIPPER MICRO, 20G X .75" 12/BOX; 56. NDL, GRIPPER MICRO, 20G X 1" 12/BOX; 57. NDL, GRIPPER MICRO, 20G X 1.25" 12/BOX; 58. NDL, GRIPPER MICRO, 19G X .75" 12/BOX; 59. NDL, GRIPPER MICRO, 19G X 1" 12/BOX; 60. NDL, GRIPPER MICRO, 19G X 1.25" 12/BOX; 61. NDL, GRIPPER MICRO, 22G X .75", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 62. NDL, GRIPPER MICRO, 22G X 1", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 63. NDL, GRIPPER MICRO, 22G X 1.25", NAC Y-SITE 12/BOX; 64. NDL, GRIPPER MICRO, 20G X .75", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 65. ND. Reason: Potential for certain GRIPPER Needles to contain an occluded or blocked needle..
- Recalling firm
- Smiths Medical ASD Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-07-08
- Terminated
- 2020-08-23
- Firm location
- Minneapolis, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GRIPPER Needles sold individually. Labeled as the following item descriptions: 1. NDL, GRIPPER, 22G X 1" (25MM), 12/BX; 2. NDL, GRIPPER, 22G X 1.25" (32MM), 12/BX; 3. NDL, GRIPPER, 20G X 1.25" (32MM), 12/BX; 4. NDL, GRIPPER, 19G X 1" (25MM), 12/BX; 5. NDL, GRIPPER, 19G X 1.25" (32MM), 12/BX; 6. NDL, GRIPPER, 22G X .75" (19MM), 12/BX; 7. NDL, GRIPPER, 19G X .75" (19MM), 12/BX; 8. NDL, GRIPPER, 20G X .625" (16MM), 12/BX; 9. NDL, GRIPPER, 22G X .625" (16MM), 12/BX; 10. NDL, GRIPPER PLUS, 22G X 1" (25MM), 12/BX; 11. NDL, GRIPPER PLUS, 22G X 1.25" (32MM), 12/BX; 12. NDL, GRIPPER PLUS, 20G X 1" (25MM) 12/BX; 13. NDL, GRIPPER PLUS, 20G X 1.25" (32MM) 12/BX; 14. NDL, GRIPPER PLUS, 19G X 1" (25MM) 12/BX; 15. NDL, GRIPPER PLUS, 19G X 1.25" (32MM) 12/BX; 16. NDL, GRIPPER PLUS, 22G X .75" (19MM) 12/BX; 17. NDL, GRIPPER PLUS, 20G X .75" (19MM) 12/BX; 18. NDL, GRIPPER PLUS, 19G X .75" (19MM) 12/BX; 19. NDL, GRIPPER PLUS, 20G X .625" (16MM) 12/BX; 20. NDL, GRIPPER PLUS, 22G X .625" (16MM) 12/BX; 21. NDL, GRIPPER PLUS, 22G X .625" (16MM), Y-SITE 12/BX; 22. NDL, GRIPPER PLUS, 22G X .75" (19MM), Y-SITE 12/BX; 23. NDL, GRIPPER PLUS, 22G X 1" (25MM), Y-SITE 12/BX; 24. NDL, GRIPPER PLUS, 22G X 1.25" (32MM), Y-SITE 12/BX; 25. NDL, GRIPPER PLUS, 20G X .625" (16MM), Y-SITE 12/BX; 26. NDL, GRIPPER PLUS, 20G X .75" (19MM), Y-SITE 12/BX; 27. NDL, GRIPPER PLUS, 20G X 1" (25MM), Y-SITE 12/BX; 28. NDL, GRIPPER PLUS, 20G X 1.25" (32MM), Y-SITE 12/BX; 29. NDL, GRIPPER PLUS, 19G X .75" (19MM), Y-SITE 12/BX; 30. NDL, GRIPPER PLUS, 19G X 1" (25MM), Y-SITE 12/BX; 31. NDL, GRIPPER PLUS, 19G X 1.25" (32MM), Y-SITE 12/BX; 32. NDL, GRIPPER, 22G X .625" (16MM), Y-SITE 12/BX; 33. NDL, GRIPPER, 22G X .75" (19MM), Y-SITE 12/BX; 34. NDL, GRIPPER, 22G X 1" (25MM), Y-SITE 12/BX; 35. NDL, GRIPPER, 22G X 1.25" (32MM), Y-SITE 12/BX; 36. NDL, GRIPPER, 20G X .625" (16MM), Y-SITE 12/BX; 37. NDL, GRIPPER, 20G X .75" (19MM), Y-SITE 12/BX; 38. NDL, GRIPPER, 20G X 1" (25MM), Y-SITE 12/BX; 39. NDL, GRIPPER, 20G X 1.25" (32MM), Y-SITE 12/BX; 40. NDL, GRIPPER, 19G X .75" (19MM), Y-SITE 12/BX; 41. NDL, GRIPPER, 19G X 1.25" (32MM), Y-SITE 12/BX; 42. NDL, GRIPPER PLUS, 22G X .625" (16MM), NEEDLELESS Y-SITE 12/BX; 43. NDL, GRIPPER PLUS, 22G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 44. NDL, GRIPPER PLUS, 22G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 45. NDL, GRIPPER PLUS, 20G X .625" (16MM), NEEDLESS Y-SITE 12/BX; 46. NDL, GRIPPER PLUS, 20G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 47. NDL, GRIPPER PLUS, 20G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 48. NDL, GRIPPER PLUS, 20G X 1.25" (32MM), NEEDLELESS Y-SITE 12/BX; 49. NDL, GRIPPER PLUS, 19G X .75" (19MM), NEEDLELESS Y-SITE 12/BX; 50. NDL, GRIPPER PLUS, 19G X 1" (25MM), NEEDLELESS Y-SITE 12/BX; 51. NDL, GRIPPER PLUS, 19G X 1.25" (32MM), NEEDLELESS Y-SITE 12/BX; 52. NDL, GRIPPER MICRO, 22G X .75 12/BOX; 53. NDL, GRIPPER MICRO, 22G X 1" 12/BOX; 54. NDL, GRIPPER MICRO, 22G X 1.25" 12/BOX; 55. NDL, GRIPPER MICRO, 20G X .75" 12/BOX; 56. NDL, GRIPPER MICRO, 20G X 1" 12/BOX; 57. NDL, GRIPPER MICRO, 20G X 1.25" 12/BOX; 58. NDL, GRIPPER MICRO, 19G X .75" 12/BOX; 59. NDL, GRIPPER MICRO, 19G X 1" 12/BOX; 60. NDL, GRIPPER MICRO, 19G X 1.25" 12/BOX; 61. NDL, GRIPPER MICRO, 22G X .75", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 62. NDL, GRIPPER MICRO, 22G X 1", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 63. NDL, GRIPPER MICRO, 22G X 1.25", NAC Y-SITE 12/BOX; 64. NDL, GRIPPER MICRO, 20G X .75", LUER ACTIVATED NEEDLELESS Y-SITE 12/BOX; 65. ND
Reason for recall
Potential for certain GRIPPER Needles to contain an occluded or blocked needle.