Recall Z-0188-2026— BALT USA, LLC
Class I · Ongoing
Class I FDA medical device recall by BALT USA, LLC, initiated 2025-09-25. It names one FDA 510(k) clearance: K242376. Product: Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component. Reason: Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen..
- Recalling firm
- BALT USA, LLC
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2025-09-25
- Firm location
- Irvine, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09180, MEGABT09190, MEGABT091100, MEGABT091110 Software Version: N/A Product Description: The Mega Ballast 091 is designed with a semi-rigid proximal shaft and a highly flexible distal shaft. The design facilitates the advancement of the Mega Ballast to access the tortuous vasculature over a guidewire and a respective dilator. The outer surface of the Mega Ballast is coated with a hydrophilic coating to facilitate lubricity from tip to 40 cm proximal. The proximal end of the Mega Ballast incorporates a Luer fitting hub for the attachment of accessories. The Mega Ballast is provided sterile, non-pyrogenic, and is intended for single use only. Component: Not a component
Reason for recall
Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.
Data sourced from openFDA. This site is unofficial and independent of the FDA.