Recall Z-0188-2015— GE Healthcare, LLC

Class II · Terminated

Class II FDA medical device recall by GE Healthcare, LLC, initiated 2014-10-14, terminated 2015-02-27. It names 3 FDA 510(k) clearances: K052434, K022960, K991841. Product: Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies. Reason: GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems..

Recalling firm
GE Healthcare, LLC
Classification
Class II
Status
Terminated
Initiated
2014-10-14
Terminated
2015-02-27
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Optima NM/CT 640, Models H3100YA, H3100YC. Product Usage: The GE Optima NM/CT 640 system is intended for General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine (NM) system using a variety of scanning modes supported by various acquisition types, and a CT component which is intended specifically for enabling attenuation correction and anatomical localization on SPECT studies

Reason for recall

GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.