Recall Z-0187-2018— Physio-Control, Inc.

Class I · Terminated

Class I FDA medical device recall by Physio-Control, Inc., initiated 2017-10-27, terminated 2019-06-12. It names one FDA 510(k) clearance: K022732. Product: LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.. Reason: The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols..

Recalling firm
Physio-Control, Inc.
Classification
Class I
Status
Terminated
Initiated
2017-10-27
Terminated
2019-06-12
Firm location
Redmond, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

LIFEPAK(R) Electrode Starter Kits, Catalog # 11101-000017, MIN #3202784-009 The defibrillation electrodes are used only with LIFEPAK EXPRESS(R) AED, LIFEPAK CR(R) Plus AED, LIFEPAK 1000 defibrillator, or LIFEPAK 500 Biphasic AED s with a pink connector.

Reason for recall

The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.