Recall Z-0186-2014— Fresenius Medical Care Holdings, Inc.
Class II · Terminated
Class II FDA medical device recall by Fresenius Medical Care Holdings, Inc., initiated 2013-08-29, terminated 2021-02-11. It names one FDA 510(k) clearance: K070049. Product: Combiset Blood Line Hemodialysis use Catalog Number: 03-2622-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.. Reason: Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss.
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-08-29
- Terminated
- 2021-02-11
- Firm location
- Waltham, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Combiset Blood Line Hemodialysis use Catalog Number: 03-2622-3. Intended for use as the extracorporeal blood circuit during Hemodialysis and is indicated for use with conventional and high flux negative pressure hemodialyzer equipment.
Reason for recall
Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss
Data sourced from openFDA. This site is unofficial and independent of the FDA.