Recall Z-0185-2026— American Contract Systems Inc.
Class II · Ongoing
Class II FDA medical device recall by American Contract Systems Inc., initiated 2025-09-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY. Reason: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed..
- Recalling firm
- American Contract Systems Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2025-09-04
- Firm location
- Grimes, IA, United States
Product description
Medical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
Reason for recall
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Data sourced from openFDA. This site is unofficial and independent of the FDA.