Recall Z-0184-2013— Wright Medical Technology Inc

Class II · Terminated

Class II FDA medical device recall by Wright Medical Technology Inc, initiated 2012-09-05, terminated 2014-01-07. It names 6 FDA 510(k) clearances: K111698, K092201, K052915 and 3 more. Product: Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI. Reason: There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility..

Recalling firm
Wright Medical Technology Inc
Classification
Class II
Status
Terminated
Initiated
2012-09-05
Terminated
2014-01-07
Firm location
Arlington, TN, United States

510(k) clearances named in this recall

6 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI

Reason for recall

There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.

Data sourced from openFDA. This site is unofficial and independent of the FDA.