Recall Z-0181-2024— American Contract Systems, Inc.

Class II · Ongoing

Class II FDA medical device recall by American Contract Systems, Inc., initiated 2023-09-01. FDA's recall record names no specific 510(k) clearance for this event. Product: Pediatric Urology Wolson Pack, REF BPPU55A. Reason: During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals..

Recalling firm
American Contract Systems, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-09-01
Firm location
Temple Terrace, FL, United States

Product description

Pediatric Urology Wolson Pack, REF BPPU55A

Reason for recall

During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.

Data sourced from openFDA. This site is unofficial and independent of the FDA.