Recall Z-0180-2022— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2021-09-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543. Reason: The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2021-09-20
Firm location
Tarrytown, NY, United States

Product description

Atellica CH Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11097543

Reason for recall

The firm has confirmed the potential for Atellica CH Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention.